Part 3, Chapter 22: Manual Appendix 3-22-A
- Machine Checkout Overview.
* = Footnote
- * Verify that backup ventilation equipment is available and functioning.
- * Check oxygen cylinder supply.
- * Check central pipeline supplies.
- * Check initial status of low pressure system.
- * Perform leak check of machine low pressure system.
- * Turn on machine master switch and all other necessary electrical equipment.
- * Test flowmeters.
- * Adjust and check scavenging system.
- * Calibrate O2 monitor.
- Check initial status of breathing system.
- Perform leak check of breathing system.
- Test ventilation systems and unidirectional valves.
- Check, calibrate and/or set alarm limits of all monitors.
- Check final status of machine.
- Procedure.
- * Emergency ventilation equipment: Verify backup ventilation equipment is available and functioning.
- * High pressure system
- *Low Pressure System.
- * Check initial status of low pressure system.
- Close flow control valves and turn off vaporizers.
- Check fill level and tighten vaporizers' fill caps.
- * Perform leak check of machine low pressure system.
- Verify that the machine master switch and flow control valves are off.
- Attach "suction bulb" to common (fresh) gas outlet.
- Squeeze bulb repeatedly until fully collapsed.
- Verify that bulb stays fully collapsed for at least 10 seconds.
- Open one vaporizer at a time and repeat 'c' and 'd' above.
- Remove suction bulb and reconnect fresh gas hose.
- * Turn on machine master switch and all other necessary electrical equipment.
- * Test flowmeters.
- Adjust flow of all gases through their full range, checking for smooth operation of floats and undamaged flow tubes.
- Attempt to create hypoxic O2 /N2O mixture and verify that correct changes in flow and/or alarm.
- * Check initial status of low pressure system.
- * Scavenging System. Adjust and check scavenging system.
- Ensure proper connections between the scavenging system and both APL (pop-off) valve and ventilator relief valve.
- Adjust gas vacuum, if possible.
- Fully open APL valve and occlude Y-piece.
- With minimum O2 flow, allow scavenger reservoir bag to collapse completely and verify that absorber pressure gauge reads about zero.
- With the O2 flush activated, allow the scavenger reservoir bag to distend fully and then verify that absorber pressure gauge reads < 10 cm H2O.
- Breathing System.
- * Calibrate O2 monitor.
- Ensure monitor reads 21% in room air.
- Verify low O2 alarm is enabled and functioning.
- Reinstall sensor in circuit and flush breathing system with O2.
- Verify that monitor now reads greater than 90%.
- Check initial status of breathing system.
- Set selector switch to "Bag" mode.
- Check that breathing circuit is complete, undamaged and unobstructed.
- Verify that CO2 absorbent is adequate.
- Install breathing circuit accessory equipment (e.g. humidifier, PEEP valve) to be used during case.
- Perform leak check of breathing system.
- Set all gas flows to zero (or minimum).
- Close APL (pop-off) valve and occlude Y-piece.
- Pressurize breathing system to about 30 cm H2O with O2 flush.
- Ensure that pressure remains fixed for at least 10 seconds.
- Open APL (pop-off) and ensure that pressure decreases.
- * Calibrate O2 monitor.
- Test manual and automatic ventilation systems and unidirectional valves.
- Place a second breathing bag on Y-piece.
- Set appropriate ventilator parameters for next patient.
- Switch to automatic ventilation (ventilator) mode.
- Turn ventilator ON and fill bellows and breathing bag with O2 flush.
- Set O2 flow to minimum, other gas flows to zero.
- Verify that during inspiration bellows delivers appropriate tidal volume and that during expiration bellows fills completely.
- Set fresh gas flow to 5 L/min.
- Verify that the ventilator bellows and simulated lungs fill and empty appropriately without sustained pressure at end expiration.
- Check for proper action of unidirectional valves.
- Exercise breathing circuit accessories to ensure proper function.
- Turn ventilator OFF and switch to manual ventilation (Bag/APL) mode.
- Ventilate manually and assure inflation and deflation of artificial lungs and appropriate feel of system resistance and compliance.
- Remove second breathing bag from Y-piece.
- Monitors. Check, calibrate and/or set alarm limits of all monitors.
- Capnometer
- Oxygen analyzer
- Pressure monitor with high and low airway pressure alarms
- Pulse oximeter
- Respiratory volume monitor (spirometer)
- Check final status of machine.
- Vaporizers off
- APL valve open
- Selector switch to "Bag"
- All flowmeters to zero (or minimum)
- Patient suction level adequate
- Breathing system ready to use
* If an anesthesia provider uses the same machine in successive cases during a day, these steps nee not be repeated or may be abbreviated after initial checkout (U.S. Food and Drug Administration, 1993).